Home > Products > Spectrometer > Echelle spectrometer

DEMON

Related Applay:

High spectral resolution

High optical throughput

Suitable with various detectors

Motorized optomechanics

High spectral resolution Double Echelle Monochromator series in Littrow configuration for the measurement of emission and absorption lines from the UV up to the NIR range.

166 000 75669HOTLINE
WeChat Service

Introduction

  • SETUP

The DEMON is a spectral high-resolution Double Echelle MONochromator. It consists of an optomechanically motorized and thermally robust echelle spectrometer in sequence with a prism monochromator that is used for the selection of the inspection range. The patented optical design of prism and echelle grating is arranged in Littrow configuration and benefits from the high image quality of the applied parabolic mirror optics. The width of the prism monochromator exit slit, which serves also as entrance slit of the echelle spectrometer, is adjustable. An active wavelength stabilization is provided by using the internal calibration lamp for highest absolute wavelength accuracy without the necessity of extensive temperature stabilization.


By applying reflection optics with broadband UV layers and a CaF2 prism, chromatic aberrations are avoided and thus a large wavelength range from the UV to the NIR is provided. The light coupling into the DEMON spectrometer is realized via SMA fiber or reflection transfer optics. Various CCD, ICCD and InGaAs cameras are suitable for the designed detection area.

The supplied operating software Sophi with optional LabVIEW library for complete remote control of the DEMON allows fully access of all spectrometer-detector functions via graphic user interface and provides automated measurement routines with an integrated scripting language. All measured spectra are automatically plotted for comprehensive analysis that can easily be performed with Sophi.

  • MODELS

ModelDEMONDEMON-NIRSUPER DEMON
Optical designechelle spectrometer with pre-monochromator and
active wavelength stabilization in Littrow-configuration
Aperturef/10 f/20
Slit widthadjustable
Wavelength range190-900 nm (more on request)600-1,700 nm200-750 nm
Spectral resolving power λ/min measurable FWHM75,000 (150,000 possible)60,000200,000
Spectral resolution2.5-12 pm10-28 pm1-3.75 pm
Absolute accuracyspectral resolution / 4pixel dispersion x 10 pixelspectral resolution / 4
Simultaneous inspection range1-5 nm4.5-13.5 nm0.25-1 nm
Linear dispersion*λ / 225,000λ / 180,000λ / 600,000
DetectorCCD or ICCDInGaAsCCD or ICCD
Exposure time, min.1 ms with CCD; 1 ms with InGaAs; 5 ns with ICCD
Dynamic range16 bit
Light couplingUV fiber down to 190 nm or mirror optics down to 175 nm
Wavelength calibrationwith calibration lamp
Controlautomatic control of the motors and the calibration lamps via PC
SoftwareSophi, LabVIEW library optional
Dimensions without detector (L x W x H)(600 x 310 x 230) mm(1215 x 512 x 300) mm
Weight without detector25 kg75 kg
Featuresintegrated mechanical shutter and motorized slit

 

* Values apply for a CCD with 1,024 pixels and a pixel size of 13.5 µm
Other spectral resolutions and wavelength ranges on request
Subject to technical changes without notice

  • APPLICATIONS