Home > Products > Spectrometer > Echelle spectrometer

ARYELLE 400

Related Applay:

High spectral resolution

Large wavelength range

High light throughput

The flagship of ARYELLE echelle spectrometer series with very high spectral resolution in the VUV or UV-VIS-NIR wide-range with 400 mm focal length – for precise elemental analysis with LIBS and Raman spectroscopy.

166 000 75669HOTLINE
WeChat Service

Introduction

  • SETUP

The ARYELLE 400 with a focal length of 400 mm and an aperture of f/10 features the highest spectral resolution of the wide-range spectrometer ARraY echELLE series. The patented optical setup with echelle grating, prism, imaging optics and entrance slit can individually be adapted to different customer applications. Depending on the required simultaneously detectable wavelength range, the ARYELLE 400 provides the measurement of two-dimensional echelle spectra. At standard spectrometer configuration with 50 µm entrance slit, a spectral resolving power of up to 30,000 for the VUV range (175-330 nm) or up to 15,000 for the UV-VIS-NIR range (330-850 nm) can be achieved. The optomechanical setup is designed to fit various CCD and ICCD cameras. Automatic recalibration of the wavelength scale is provided by a mercury lamp and an internal shutter.


The supplied operating software Sophi with optional LabVIEW library allows fully control of all spectrometer-detector functions via graphic user interface and provides automated measurement routines with an integrated scripting language. All lines of the gained spectrum are automatically analyzed with the integrated atomic spectra data base and qualitatively assigned to the corresponding elements. Quantitative analysis is based on calibration with reference materials.

  • SPECIFICATIONS

Optical designechelle spectrograph
Aperturef/10
Focal length400 mm
Slit width50 µm
Absolute accuracyspectral resolution / 4
Dynamic range15 bit, A/D conversion 16 bit
Light couplingSMA fiber coupling or mirror optics
Wavelength calibrationwith mercury lamp
ComputerPC incl. TFT or laptop with Microsoft Windows
SoftwareSophi, LabVIEW library optional
Dimensions without detector (L x W x H)(438 x 200 x 232) mm
Weight without detector12 kg

 

Setup123456
DetectorCCDCCDCCDCCDICCDICCD
Wavelength range1190-330 nm330-850 nm192-930 nm250-900 nm192-433 nm425-750 nm
Spectral resolving power
λ/min. measurable FWHM1
30,00015,00015,00045,00014,00020,000
Spectral resolution1,26-11 pm22-57 pm13-62 pm5.6-20 pm13-31 pm21-37 pm
Detector resolution2,048 x 512 px2,048 x 512 px2,048 x 2,048 px2,048 x 2,048 px1,024 x 1,024 px1,024 x 1,024 px
Image area27.6 x 6.9 mm²27.6 x 6.9 mm²27.6 x 27.6 mm²27.6 x 27.6 mm²13.3 x 13.3 mm²13.3 x 13.3 mm²
Chopperyesyesyesyesnono
Step width, min.0.1 µs0.1 µs0.1 µs0.1 µs1 ns1 ns
Gate width5 ns5 ns

 

1 Depending on the chosen grating and prism, other ranges possible (down to 175 nm)
2 Depending on the slit width, other widths possible

Other spectral resolutions, wavelength ranges, CCD and ICCD detectors on request