Home > Products > Spectrometer > Mid-infrared spectrometer

OEM FT-IR module

Related Applay:

All-in-one FT-IR

Compact and rugged

High resolution 

Cost effective

The ARCoptix OEM FT-IR module is the ideal product for system integrators (or labatoratories) looking for a customized FTIR measurement design. 

166 000 75669HOTLINE
WeChat Service

Introduction

ARCoptix OEM FT-IR module

Robust, compact and reliable OEM FT-IR module optimized for system integration

The ARCoptix OEM FT-IR module is the ideal product for system integrators (or labatoratories) looking for a customized FTIR measurement design. The OEM module exists in two configurations:

  • Fully integrated OEM module (OEM010) (as shown on picture) with integrated light source and detector.
  • Flexible OEM module (OEM011) with external detector module.


The ARCoptix OEM010 is our most compact FT-IR complete system. It features a built-in SiC globar heated to 1550 °K and regulated in temperature to ensure stable illumination. The light that is being output by the instrument is first modulated by our permanently aligned interferometer, making it robust against stray light that might eventually reach the detector. The OEM010 rugged casing also holds a highly sensitive TE-MCT detector which maximizes the response of the system. Built from the same components, the OEM011 displays identical performances as the OEM010 while providing with an added degree of freedom thanks to its separate detector module.


  • Characteristics

  • uAll-in-one FT-IR or flexible confiuration

  • uCompact and rugged

  • uHigh resolution of 2 cm-1 (optionally up to 0.5 cm-1)

  • uEasily connected to IR fiber systems

  • u3 possible TE-cooled MCT detector ranges:5000-1700cm-1 (2-6μm), 6700-1200cm-1 (1.5 - 8.5μm) or 5000-830 cm-1 (2-12μm)

  • uUSB connectivity

  • uCost effective


  • Specifications

Product codeFTIR-OEM01X-060-4TE
FTIR-OEM01X-085-4TE
FTIR-OEM01X-120-4TE
Beam-spliter materialCaF2ZnSe
Spectral range [cm-1]5'000 - 1'6605'000 - 1'2005'000 - 830
Spectral range [μm]2.0-6.01.5-8.52.0-12.0
D* [cm Hz½W-1μm]>1x10E11>8x10E10>4x10E9
Signal-to-Noise (4cm-1 resolution, 1 min scan time)> 80'000:1> 40'000:1


 Common specifications

Detector type

4 stage TE cooled MCT detector

Interferometer type

Permanently aligned, double retro-reflector design

Resolution [cm-1 ]
(unapodized)

8, 4, 2, optionally 1 and 0.5

Wavenumber repetability

<10ppm
Scan frequency

>4 Hz @ 4cm-1

Control laser

Temperature-stabilized
solid-state laser @850nm

A/D Converter24 bit
Operating temperature10°C-40°C
Free-space inputAperture Ø: 12.7mm
Acceptance full angle: 3.2°
Removable fiber-couplersConnector: SMA 905
Fiber core Ø: up to 1mm
NA=0.25 (lensed coupler) or NA=0.3 (reflective coupler)
Power requirement12V / 40W
Communication InterfaceUSB 2.0
Software Interface

Windows 7/10/11
API for controlling the instrument via our DLL

Dimensions180mm x 160mm x 80mm
Weight1800 g