Home > Products > Spectrometer > Mid-infrared spectrometer

FT-IR

Related Applay:

fibered probes

Most compact fiber-coupled FT-IR

Available with TE-cooled

The ARCoptix FT-IR spectrometer is an all-fibered alternative to our "Rocket" configurations. 

166 000 75669HOTLINE
WeChat Service

Introduction

ARCoptix fiber coupler FT-IR

The ARCoptix FT-IR spectrometer is an all-fibered alternative to our "Rocket" configurations. It features the same components as the FT-NIR or FT-MIR Rocket instruments, with the addition of an internally mounted optical source that enables to modulate light before coupling it to the output port of the device. This architecture is thus more robust to external perturbations such as background thermal emission, as such parasitic light would not go through the interferometer before reaching the detector. Our FTIR-FC is ideal for fibered applications such as reflection probes used for diffuse reflection measurements.

4-TE or Liquid-Nitrogen cooled MCT detectors available

The Arcoptix FTIR-FC can be equipped with both 4-stage thermo-electrically (TE) cooled or liquid-nitrogen cooled MCT detectors. The TE-cooled detectors in the FTIR-FC are the most sensitive TE-cooled detectors availabe on the market and offer a reliable and maintenance-free operation. For ultimate sensitivity levels, the FTIR-FC can also be equipped with a liquid-nitrogen cooled detector. In the NIR, the FTIR-FC can also feature an extended InGaAs detector.


  • Characteristics

  • uIdeal for operation with fibered probes

  • uAll-in-one solution: source, interferometer, detector and fiber-couplers

  • uMost compact fiber-coupled FT-IR on the market

  • uAvailable with TE-cooled or liquid-nitrogen cooled MCT detectors


  • Specifications

Product codeFTNIR-FC-025-2TE
FTMIR-FC-060-4TE
FTMIR-FC-120-4TE
FTMIR-FC-160-LN2
Beam-spliter materialCaF2ZnSe
Spectral Range [cm-1]11'000 - 4'0005'000 - 1'6605'000 - 8305000 - 650
Spectral Range [μm]0.9-2.52.0-6.02.0-12.02.0-16.0
Detector TypeInGaAs (2-TE cooled)MCT (4-TE cooled)MCT (LN2 cooled)
D* [cm Hz1/2W-1]>2x10E11>1x10E11>4x10E9>5x10E10
Signal-to-Noise (4cm-1 resolution, 1 min scan time)> 100'000:1> 40'000:1> 70'000:1
Recommended fiberLow-OH, multimode silica fiberCIR or PIR depending of the applicationPIR (polycrystalline) fiber
Internal reference laserTemperature-stabilized, @795nmTemperature-stabilized, @850nm
Built-in light source20W QTH bulb (2850°K)20 SiC globar (1550°K)
Power requirement40W @12VDC30W @12VDC

 

  • Common specifications

Interferometer type

Permanently aligned, double retro-reflector design

Resolution [cm-1 ]
(unapodized)

8,4,2

Wavenumber repetability

<10ppm
Scan frequency

>4 Hz @ 4cm-1

Control laser

Temperature-stabilized
solid-state laser @850nm

A/D Converter24 bit
Operating temperature10°C-40°C
Free-space inputAperture Ø: 12.7mm
Acceptance full angle: 3.2°
Removable fiber-couplersConnector: SMA 905
Fiber core Ø: up to 1mm
NA=0.25 (lensed coupler) or NA=0.3 (reflective coupler)
Communication InterfaceUSB 2.0
Software Interface

Windows 7/10/11
API for controlling the instrument via our DLL

Dimensions180mm x 160mm x 80mm (excluding LN2 dewar)